Z-1298-2022 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- July 6, 2022
- Initiation Date
- April 19, 2022
- Termination Date
- March 27, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 111
Product Description
Product Name: TEG5000 Analyzer (07-022, 07-033) with TEG Analytical Software (07-030, 07-031) and TEG Platelet Mapping Assay - PlateletMapping ADP & AA (07-014), PlateletMapping ADP (07-015), PlateletMapping AA (07-016). Issue occurs when Platelet Mapping Assay results from a TEG 5000 analyzer are viewed using TEG Manager Software (versions 1.1-4.3.1)
Reason for Recall
When the TEG 5000 Analyzer including TEG Analytical Software is used with PlateletMapping (ADP or AA) Assays and TEG Mangement Software, incorrect values of the PlateletMapping ADP or AA % Inhibition and % Aggretagion may be displayed on TEG Manager for the affected assays.
Distribution Pattern
US Nationwide.
Code Information
This correction is applicable when TEG 5000 PlateletMapping (Item # 07-014, 07-015, 07-016) is performed on a TEG 5000 analyzer (Item # 07-022, 07-033) that is connected to TEG Manager Software which includes viewing of TEG 5000 results. Code Info: UDI-DI: (01)30812747018184 Software Version: TEGM 1.1 and later, TAS v4.2.3