Z-0898-2020 Class II Terminated

Recalled by Haemonetics Corporation — Braintree, MA

Recall Details

Product Type
Devices
Report Date
February 5, 2020
Initiation Date
October 21, 2019
Termination Date
November 6, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
28462

Product Description

JAPAN BASIC HI SPEED CS5 SET,LN 263J - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. The recovered blood is processed by a centrifugation and washing procedure before pumping the red cell product to either a bag for gravity reinfusion into the patient or an extracorporeal circuit.

Reason for Recall

Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.

Code Information

Catalog number/FG Item Number: 0263J-00 1. Specific Lot with Bowl Cracks (814 Distributed) Lot number: 0718027 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019.