Z-1303-2018 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-1303-2018 was initiated by Siemens Medical Solutions USA, Inc on Jan 8, 2018 and is designated Class II. Reason for recall: A software functionality in the report sections Findings Information and Summary of Measured Findings is not performing properly. The system will use the original values instead of displaying and… The recall status is terminated (terminated Apr 22, 2020). Affected quantity: 236 units.

Recall Details

Product Type
Devices
Report Date
Apr 11, 2018
Initiation Date
Jan 8, 2018
Termination Date
Apr 22, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
236 units

Product Description

Syngo.via. Medical Device Software. Picture archiving and communications system.

Reason for Recall

A software functionality in the report sections Findings Information and Summary of Measured Findings is not performing properly. The system will use the original values instead of displaying and saving corrected/modified values. Outdated information could be sent to the referring physician when creating the formal report in the reporting system.

Distribution Pattern

Device is software only. No products are distributed to wholesale dealers, distributers or retailers

Code Information

Model Number 10496180

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1303-2018
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.