Z-1304-2024 Class II Ongoing

Recalled by Siemens Healthcare Diagnostics, Inc. — Tarrytown, NY

FDA device recall Z-1304-2024 was initiated by Siemens Healthcare Diagnostics, Inc. on Feb 7, 2024 and is designated Class II. Reason for recall: The reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test… The recall status is ongoing. Affected quantity: 74,211 Kits (Expanded 2/28/24: additional 72,892 kits).

Recall Details

Product Type
Devices
Report Date
Mar 27, 2024
Initiation Date
Feb 7, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
74,211 Kits (Expanded 2/28/24: additional 72,892 kits)

Product Description

DCA Systems HbA1c Reagent Kit

Reason for Recall

The reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test results could be reported and potentially lead to incorrect patient management of glucose control. A positive bias may be observed with eleven HbA1c affected lots. Positive bias was observed when the performance of these five HbA1c lots was compared to NGSP pooled patient target-value assigned samples and the maximum bias observed was 19%.

Distribution Pattern

Worldwide distribution - US Nationwide and the country of Taiwan.

Code Information

Siemens Material Number: 10311134; UDI/DI: 00630414532806; Lot#: 9640093, 0645103, 0648103, 0649103, 0654103 . Expanded Lots (02/28/2024): 0637093, 0641093, 0642093, 0647103, 0653103, 0655103;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1304-2024
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.