Z-1306-2022 Class II Ongoing

Recalled by Baxter Healthcare Corporation — Deerfield, IL

FDA device recall Z-1306-2022 was initiated by Baxter Healthcare Corporation on May 31, 2022 and is designated Class II. Reason for recall: An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through… The recall status is ongoing. Affected quantity: 283 installations.

Recall Details

Product Type
Devices
Report Date
Jul 6, 2022
Initiation Date
May 31, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
283 installations

Product Description

NaviCare Nurse Call/Voalte Nurse Call

Reason for Recall

An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority.

Distribution Pattern

nationwide and Canada

Code Information

software versions: 3.9.100, 3.9.101,3.9.102, 3.9.200, 3.9.201, and 3.9.300

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1306-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.