Z-1310-2024 Class II Ongoing

Recalled by Mc3 Inc — Dexter, MI

FDA device recall Z-1310-2024 was initiated by Mc3 Inc on Oct 11, 2023 and is designated Class II. Reason for recall: May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be… The recall status is ongoing. Affected quantity: 188 units.

Recall Details

Product Type
Devices
Report Date
Mar 27, 2024
Initiation Date
Oct 11, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
188 units

Product Description

NautilusTM Smart ECMO Module with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Product Number: 48135, 48135E

Reason for Recall

May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer

Distribution Pattern

US Nationwide distribution.

Code Information

GTIN:(1)10854916006871; (2)10854916006895 All Serial Numbers between 481022923 and 481024039

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1310-2024
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.