Z-1310-2024 Class II Ongoing

Recalled by Mc3 Inc — Dexter, MI

Recall Details

Product Type
Devices
Report Date
March 27, 2024
Initiation Date
October 11, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
188 units

Product Description

NautilusTM Smart ECMO Module with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Product Number: 48135, 48135E

Reason for Recall

May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer

Distribution Pattern

US Nationwide distribution.

Code Information

GTIN:(1)10854916006871; (2)10854916006895 All Serial Numbers between 481022923 and 481024039