Z-1312-2023 Class II Ongoing

Recalled by Defibtech, LLC — Guilford, CT

Recall Details

Product Type
Devices
Report Date
April 5, 2023
Initiation Date
February 14, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
29 units

Product Description

DDU-100 Lifeline Automated External Defibrillator (AED)-Indicated for use on victims of sudden cardiac arrest

Reason for Recall

A component of a sub-assembly used in the affected AEDs had not undergone durability and reliability testing, and if a malfunction occurs, therefore the device could fail to deliver therapy.

Distribution Pattern

US Nationwide distribution.

Code Information

UDI-DI: 00815098020003 Serial Numbers: 101702198 101702263 101702424 101702524 101702535 101702540 101702544 101702545 101703267 101703269 101703272 101703274 101703276 101703277 101703278 101703281 101703312 101703326 101703378 101703407 101703411 101703412 101703419 101703424 101703427 101703464 101703465 101703600 101704106