Z-1312-2024 Class II Ongoing

Recalled by Mckesson Medical-Surgical Inc. Corporate Office — Richmond, VA

FDA device recall Z-1312-2024 was initiated by Mckesson Medical-Surgical Inc. Corporate Office on Jan 9, 2024 and is designated Class II. Reason for recall: Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2… The recall status is ongoing. Affected quantity: 3 units.

Recall Details

Product Type
Devices
Report Date
Mar 27, 2024
Initiation Date
Jan 9, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3 units

Product Description

Abbott Rapid Dx N America LLC Reproductive Endocrinology Control Set McKesson Consult" Pregnancy (hCG) Testing Positive Level / Negative Level 2 X 4 mL

Reason for Recall

Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.

Distribution Pattern

US Nationwide distribution in the states of AZ, PA and TX.

Code Information

Catalogue Number: 5011; UDI/DI: 612479202788; Lot No. KN00589;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1312-2024
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.