Z-1326-2020 Class III Terminated

Recalled by Theralase Inc. — East York

Recall Details

Product Type
Devices
Report Date
February 26, 2020
Initiation Date
September 26, 2019
Termination Date
December 3, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
76 units

Product Description

TLC-2000 Therapeutic Medical Laser System - Product Usage: Indicated for the temporary relief of knee pain.

Reason for Recall

Power supply (Model Number: TLC-2003) ME90A2403F01 (24 VDC, 3.75 A, 90 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 4th edition specification) shipped versus Power supply (Model Number: TLC-2003) MENB1100A2403F01 (24 VDC, 4.2 A, 101 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 3rd edition specification)

Distribution Pattern

Worldwide distribution - US Nationwide and countries Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania.

Code Information

Serial Numbers: 3275 to 3350 (inclusive)