Z-1348-2020 Class III Terminated

Recalled by Theralase Inc. — East York, N/A

Recall Details

Product Type
Devices
Report Date
March 4, 2020
Initiation Date
September 26, 2019
Termination Date
December 3, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
202 units

Product Description

TLC-2000 Therapeutic Medical Laser System Power Pack, Model Number TLC-2002

Reason for Recall

Power Pack (component) was shipped with an internal battery pack tested to UN-38.3, UL-2054 and IEC-62133 (2nd Edition) standards; however, not tested to IEC-60601-1-2 (3rd edition) standards

Distribution Pattern

U.S.A.; Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania

Code Information

Serial Numbers: 3104-3106 (inclusive); 3109 to 3141 (inclusive), 3143, 3146, 3147, 3155, 3158, 3159, 3163-3165 (inclusive), 3168-3173 (inclusive), 3180-3184 (inclusive), 3188-3332 (inclusive), 3334