Z-1330-2024 Class II Ongoing

Recalled by Covidien — Lafayette, CO

FDA device recall Z-1330-2024 was initiated by Covidien on Feb 21, 2024 and is designated Class II. Reason for recall: Packs including 5 laparotomy sponges may contain incorrect product with a blue loop, X-ray detectable feature, and ethylene oxide treatment; however, this incorrect product wasn't steam pre-treated… The recall status is ongoing. Affected quantity: 755,800.

Recall Details

Product Type
Devices
Report Date
Mar 27, 2024
Initiation Date
Feb 21, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
755,800

Product Description

Situate Laparotomy Sponge, RF and X-ray Detectable, REF: L1818-04P01C-1

Reason for Recall

Packs including 5 laparotomy sponges may contain incorrect product with a blue loop, X-ray detectable feature, and ethylene oxide treatment; however, this incorrect product wasn't steam pre-treated prior to sterilization, and has no RF tag, which if non-sterile, may lead to infection, tissue trauma, and sepsis, and if in the scannable area, would lead a "CLEAR" display.

Distribution Pattern

US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI, WV.

Code Information

UDI-DI: 20884521700953, 10884521700956. Lots: 230504KF, 220903KF, 221102KF

Source