Z-1334-2024 Class II Ongoing

Recalled by Angiodynamics, Inc. — Queensbury, NY

FDA device recall Z-1334-2024 was initiated by Angiodynamics, Inc. on Jan 30, 2024 and is designated Class II. Reason for recall: Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of… The recall status is ongoing. Affected quantity: 104 units.

Recall Details

Product Type
Devices
Report Date
Mar 27, 2024
Initiation Date
Jan 30, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
104 units

Product Description

MINI STICK MAX 4F X 10 CM STD .018 SS/PD ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-749

Reason for Recall

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.

Code Information

UDI: 15051684022927 UPN: H965457491 Lot Number: 5805056

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1334-2024
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.