Z-1343-2022 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- July 13, 2022
- Initiation Date
- May 24, 2022
- Termination Date
- January 22, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 233 worldwide; 32 US
Product Description
Artis zeego (Model no. 10280959) and Artis Q.zeego (Model no. 10848283), Interventional Fluoroscopic X-Ray System
Reason for Recall
It may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are blocked and can only be reactivated by a field service engineer.
Distribution Pattern
Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CO, FL, IL, LA, ME, MN, MO, NC, OH, PA, SC, TN, TX, WI and the countries of Australia, Austria, Azerbaijan, Belgium, Canada, China, Finland, France, Germany, Hong Kong, India, Italy, Japan, Lebanon, Lithuania, Luxembourg, Netherlands, Poland, Portugal, South Korea, Switzerland, United Kingdom.
Code Information
UDI-DI: Artis zeego 04056869010083 Artis Q.zeego 04056869010007