Z-1357-2018 Class II Terminated

Recalled by C.R. Bard, Inc. — Covington, GA

Recall Details

Product Type
Devices
Report Date
April 18, 2018
Initiation Date
February 27, 2018
Termination Date
July 11, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4800 units

Product Description

Bard(R) Premature Infant Feeding tube, ENFit(TM), REF EN0036430 Product Usage: BARD(R) Nasogastric Sump Tubes with ENFit(TM) connector are intended to be used for: Decompression of stomach by suction or aspiration of gastric contents Short-term administration of term tube feeding, lavage fluid and medications Feeding Tubes with the ENFit(TM) Connector are made of a smooth plastic with a two-eyed tube used for nasogastric feeding of neonates, infants and adults as well as diagnostic and therapeutic aspiration. These products are sold single-use, sterile. The Feeding Tube with the ENFit(TM) connector provides a way to reduce the risk of enteral device misconnections. The ENFit(TM) connector: Addresses patient side connections between feeding tubes, administration sets, medication, flush and bolus feeding syringes, and other enteral devices. Does not allow connectivity with any other connector for any other clinical use.

Reason for Recall

BMD has identified that the product code/lot number combination identified above may have ineffectively fused tubing and ENFit(TM) connector.

Distribution Pattern

US Nationwide Distribution

Code Information

Lot Numbers: NGBN2862, NGAXX217, NGAVX225