Z-1370-2016 Class II Terminated

Recalled by Boston Scientific Corporation — Maple Grove, MN

FDA device recall Z-1370-2016 was initiated by Boston Scientific Corporation on Mar 28, 2016 and is designated Class II. Reason for recall: Boston Scientific is voluntarily implementing a Medical Device Recall of one batch of PT Graphix Magnet J-TIP Guidewires. Boston Scientific has found that, while the batch is labeled as a J-tip… The recall status is terminated (terminated Nov 7, 2016). Affected quantity: 47 units.

Recall Details

Product Type
Devices
Report Date
Apr 20, 2016
Initiation Date
Mar 28, 2016
Termination Date
Nov 7, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
47 units

Product Description

Boston Scientific, PT Graphix Magnet J-TIP Guidewire with ICE Hydrophilic Coating, Sterile EO, UPN H7491490201MJ2 (Carton Label) ,H7491490201MJo (UPN referenced on the Pouch Label) . Boston Scientific PT Graphix Magnet Guidewire with ICE Hydrophilic Coating is a steerable guidewire available with a nominal diameter of 0.014 in (0.37 mm) and in nominal lengths of 182 cm. PT Graphix Magnet is available with a shapeable Straight Tip or a preformed "J"¿ Tip. The PT Graphix Magnet Guidewire features a 10 or 38 cm polymer sleeve, coated with ICE Hydrophilic Coating, jacketing the distal tapered core wire. The proximal section of the PT Graphix Magnet Guidewire is coated with a fluorinated polymer.

Reason for Recall

Boston Scientific is voluntarily implementing a Medical Device Recall of one batch of PT Graphix Magnet J-TIP Guidewires. Boston Scientific has found that, while the batch is labeled as a J-tip model, the guidewires are straight and do not have the pre-formed J-tip. The most severe injury expected to occur is a limited procedure delay caused by the need to replace the guidewire with one that has

Distribution Pattern

US nationwide in the states of ME, LA, and TN

Code Information

Lot: 17971870

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1370-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.