Z-1370-2024 Class II Ongoing
FDA device recall Z-1370-2024 was initiated by Alcon Research, LLC on Feb 21, 2024 and is designated Class II. Reason for recall: Medline Bulb irrigation syringes recalled due to potential for packaging sterile barrier to be breached, and these syringes were packed inside surgical procedure packs that were subsequently… The recall status is ongoing. Affected quantity: 886.
Recall Details
- Product Type
- Devices
- Report Date
- Apr 3, 2024
- Initiation Date
- Feb 21, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 886
Product Description
SYRINGE,BULB,2OZ/60CC,GREEN, Part: 300055070, packaged inside Alcon Custom-Pak Surgical Procedure Packs, Products: 18018-06, 3865-44, 5675-76, 13554-11, 18018-06, 15190-16 , 15190-16, 15190-16, 15190-16, 15190-16, 13554-11, 5918-39, 15190-16, 15190-16, 9208-66, 9208-66, 5675-76, 5918-39, 9208-66, 15190-16, 9065-25, 5918-39, 16066-15, 16066-15
Reason for Recall
Medline Bulb irrigation syringes recalled due to potential for packaging sterile barrier to be breached, and these syringes were packed inside surgical procedure packs that were subsequently terminally sterilized, so after the packs are opened, syringes should be discarded at surgical case end, and not reshelved, because the use of non-sterile bulb irrigation syringes may increase infection risk.
Distribution Pattern
US Nationwide distribution in the states of OH, WV.
Code Information
Product/Lot/Expiration/UDI-DI: 18018-06/15T707/02/28/2025/380655148440, 3865-44/15T2RC/02/28/2025/380655148150, 5675-76/15T70E/02/28/2025/380655148174, 13554-11/15V4RC/02/28/2025/380655143711, 18018-06/15V3E0/02/28/2025/380655148440, 15190-16/JZ5179517/06/30/2024/380655148303, 15190-16/JZ5179519/06/30/2024/380655148303, 15190-16/JZ5179521/06/30/2024/380655148303, 15190-16/JZ5179523/06/30/2024/380655148303, 15190-16/JZ5179529/06/30/2024/380655148303, 13554-11/15TAUD/01/31/2026/380655143711, 5918-39/15TE5F/02/28/2025/380655149140, 15190-16/15THVA/04/30/2025/380655148303, 15190-16/15TNXJ/04/30/2025/380655148303, 9208-66/15V8CU/04/30/2025/380651693326, 9208-66/15XVDD/05/31/2025/380651693326, 5675-76/JZ5180811/06/30/2024/380655148174, 5918-39/JZ5180812/06/30/2024/380655149140, 9208-66/JZ5180332/03/31/2025/380651693326, 15190-16/JZ5180856/05/31/2025/380655148303, 9065-25/JZ5187593/11/30/2025/380651682849, 5918-39/JZ5187260/06/30/2024/380655149140, 16066-15/JZ5190913/07/31/2025/380655149683, 16066-15/JZ5191891/07/31/2025/380655149683 Bulb Syringe lots: 96922110001 or 96923100001
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1370-2024
- openFDA file date
- Oct 26, 2026
- Loaded by RegDataLab