Z-1376-2013 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-1376-2013 was initiated by Zimmer, Inc. on Mar 27, 2013 and is designated Class II. Reason for recall: Zimmer is initiating a lot specific recall of the Trabecular Metal" Tibia Stop Drill due to the potential of the outside diameter being oversized. As a result, there is a potential for the drill to… The recall status is terminated (terminated Jan 26, 2017). Affected quantity: 69 units.

Recall Details

Product Type
Devices
Report Date
Jun 5, 2013
Initiation Date
Mar 27, 2013
Termination Date
Jan 26, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
69 units

Product Description

NEXGEN Complete Knee Solution Monoblock Tibial Drill with stop 10.7 mm. Used to prepare cylindrical holes in the native tibia.

Reason for Recall

Zimmer is initiating a lot specific recall of the Trabecular Metal" Tibia Stop Drill due to the potential of the outside diameter being oversized. As a result, there is a potential for the drill to not pass through the applicable drill guide.

Distribution Pattern

Nationwide Distribution.

Code Information

Part No. 00-5887-052-00; lots 62181290 and 62183430