Z-1377-2018 Class II Terminated

Recalled by Cook Inc. — Bloomington, IN

Recall Details

Product Type
Devices
Report Date
April 25, 2018
Initiation Date
January 12, 2018
Termination Date
May 8, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
483

Product Description

MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000 US) G46022 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.

Reason for Recall

The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.

Distribution Pattern

Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide)

Code Information

All lots