Z-1378-2018 Class II Terminated

Recalled by Cook Inc. — Bloomington, IN

FDA device recall Z-1378-2018 was initiated by Cook Inc. on Jan 12, 2018 and is designated Class II. Reason for recall: The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater. The recall status is terminated (terminated May 8, 2020). Affected quantity: 4,451.

Recall Details

Product Type
Devices
Report Date
Apr 25, 2018
Initiation Date
Jan 12, 2018
Termination Date
May 8, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4,451

Product Description

MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.

Reason for Recall

The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.

Distribution Pattern

Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide)

Code Information

All Lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1378-2018
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.