Z-1388-2025 Class II Ongoing

Recalled by Beckman Coulter, Inc. — Chaska, MN

FDA device recall Z-1388-2025 was initiated by Beckman Coulter, Inc. on Feb 5, 2025 and is designated Class II. Reason for recall: the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient… The recall status is ongoing. Affected quantity: 206 units.

Recall details

Product Type
Devices
Report Date
Mar 26, 2025
Initiation Date
Feb 5, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
206 units

Product description

DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137.

Reason for recall

the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results.

Distribution pattern

US Nationwide. Global Distribution.

Code information

UDI-DI: 15099590732103; Serial Numbers: 300116-300117, 300123-300124, 300126, 300128-300139, 300141-300145, 300147- 300158, 300160-300164, 300166-300179, 300181-300183, 300186, 300188-300195, 300197-300207, 300209-300250, 300252- 300256, 300258-300259, 300261, 300263- 300264, 300266-300277, 300279-300290, 300291, 300293-300299, 300300-300309, 300312-300352

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1388-2025
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1388-2025. Report a data issue: hello@regdatalab.com.