Z-1388-2025 Class II Ongoing

Recalled by Beckman Coulter, Inc. — Chaska, MN

Recall Details

Product Type
Devices
Report Date
March 26, 2025
Initiation Date
February 5, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
206 units

Product Description

DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137.

Reason for Recall

the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results.

Distribution Pattern

US Nationwide. Global Distribution.

Code Information

UDI-DI: 15099590732103; Serial Numbers: 300116-300117, 300123-300124, 300126, 300128-300139, 300141-300145, 300147- 300158, 300160-300164, 300166-300179, 300181-300183, 300186, 300188-300195, 300197-300207, 300209-300250, 300252- 300256, 300258-300259, 300261, 300263- 300264, 300266-300277, 300279-300290, 300291, 300293-300299, 300300-300309, 300312-300352