Z-1392-2022 Class III Ongoing

Recalled by Boston Scientific Corporation — Marlborough, MA

FDA device recall Z-1392-2022 was initiated by Boston Scientific Corporation on May 31, 2022 and is designated Class III. Reason for recall: Several complaints were received for the affected lot; users received a "Fiber type does not match card type" error message on the GreenLight XPS Laser console, necessitating the replacement of the… The recall status is ongoing. Affected quantity: 57.

Recall Details

Product Type
Devices
Report Date
Jul 27, 2022
Initiation Date
May 31, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
57

Product Description

GreenLight HPSEA Laser Fibers, UPN 0010-2092

Reason for Recall

Several complaints were received for the affected lot; users received a "Fiber type does not match card type" error message on the GreenLight XPS Laser console, necessitating the replacement of the laser fiber and fiber card. This may result in prolonged procedure while the fiber is being replaced.

Distribution Pattern

Domestic distribution to AR CA IL ME MI NC TX Foreign distribution to Australia, Canada, and France.

Code Information

UDI-DI (GTIN): 00878953005522 Lot: 27770466

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1392-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.