Z-1397-2018 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-1397-2018 was initiated by Siemens Medical Solutions USA, Inc on Feb 14, 2018 and is designated Class II. Reason for recall: Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table. The recall status is terminated (terminated May 16, 2019). Affected quantity: 18.

Recall Details

Product Type
Devices
Report Date
Apr 25, 2018
Initiation Date
Feb 14, 2018
Termination Date
May 16, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
18

Product Description

MEVATRON M2 / PRIMUS Mid-Energy, Digital Linear Accelerator, Model No. 01940035 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.

Reason for Recall

Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

Distribution Pattern

Worldwide Distribution- US Nationwide in the states: AL, CA, FL, GA, IL, IN, KY, MA, MI, MO, MS, NJ, NY, OH, OK, PA, RI, SC, TN, TX, WI, WV, and WY; and PR. International distribution to Nassau.

Code Information

Serial no: 3383 5474 3296 3591 3618 3103 3986 3629 3136 70-4323 5092 3639 3274 2855 3644 4018 70-4347 3939

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1397-2018
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.