Z-1399-2024 Class II Ongoing
FDA device recall Z-1399-2024 was initiated by Becton Dickinson & Co. on Feb 29, 2024 and is designated Class II. Reason for recall: Esculin, a substrate used in other panels and normally in A16 well, was found in A17 well. A17 in yeast panels is supposed to be an empty well and is not a part of the normal ID evaluation. However… The recall status is ongoing. Affected quantity: 6700 eaches.
Recall Details
- Product Type
- Devices
- Report Date
- Apr 10, 2024
- Initiation Date
- Feb 29, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6700 eaches
Product Description
BD Phoenix Yeast ID Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyrene panel (tray) containing control wells and substrate wells with dried biochemical substrates that use traditional qualitative microbiology methods such as fermentation, oxidation, degradation, and hydrolysis in combination with chromogenic and fluorogenic substrates along with carbon sources for yeast identification.
Reason for Recall
Esculin, a substrate used in other panels and normally in A16 well, was found in A17 well. A17 in yeast panels is supposed to be an empty well and is not a part of the normal ID evaluation. However, A17 is checked for fluorescence and if anything is found, the system will abort the panel with the message Fluorescent interference has been detected. Esculin is expected to trigger this fluorescence and test cancellation.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of AR, BE, CO, IN, JP, PE, PY, SG, TW, UY.
Code Information
UDI-DI: (10) 3312180 (17) 241030 (30)0025 (01) 30382904483167 Lot Number: 3312180 Exp. Date: 30-OCT-2024