Z-1402-2018 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-1402-2018 was initiated by Siemens Medical Solutions USA, Inc on Feb 14, 2018 and is designated Class II. Reason for recall: Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table. The recall status is terminated (terminated May 16, 2019). Affected quantity: 18.

Recall Details

Product Type
Devices
Report Date
Apr 25, 2018
Initiation Date
Feb 14, 2018
Termination Date
May 16, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
18

Product Description

PRIMUS HI, Digital Linear Accelerator, Model No. 04504200 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.

Reason for Recall

Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.

Distribution Pattern

Worldwide Distribution- US Nationwide in the states: AL, CA, FL, GA, IL, IN, KY, MA, MI, MO, MS, NJ, NY, OH, OK, PA, RI, SC, TN, TX, WI, WV, and WY; and PR. International distribution to Nassau.

Code Information

Serial no: 2958 3832 3241 3574 3519 3922 3924 3361 4036 3302 70-4185 3949 3547 3657 3622 3731 3565 5153

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1402-2018