Z-1439-2017 Class II Terminated

Recalled by Smiths Medical ASD, Inc. — Saint Paul, MN

FDA device recall Z-1439-2017 was initiated by Smiths Medical ASD, Inc. on Oct 31, 2016 and is designated Class II. Reason for recall: 20 Pumps sold to the Finnish market contain a message in which one word in the message is mistranslated. When the user follows a specific set of key presses the pump will display the incorrect… The recall status is terminated (terminated Jul 15, 2019). Affected quantity: 20.

Recall Details

Product Type
Devices
Report Date
Mar 22, 2017
Initiation Date
Oct 31, 2016
Termination Date
Jul 15, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
20

Product Description

CADD Solis VIP Ambulatory Infusion Pump, Model 21-21210, Reorder 21-2120-0102-15,

Reason for Recall

20 Pumps sold to the Finnish market contain a message in which one word in the message is mistranslated. When the user follows a specific set of key presses the pump will display the incorrect message. The message indicates that a Patient Controlled Analgesia (PCA) dose is unavailable because the pump is running. It should indicate that the PCA dose is not available because the pump is stopped. The function of the pump is unchanged and no patient injury can occur since no drug is being delivered.

Distribution Pattern

Internationally to Finland

Code Information

1071155, 1076826, 1081258, 1081260, 1081275, 1082475, 1082511, 1082512, 1090347, 1090348, 1090349, 1091767, 1091768, 1091769, 1091770, 1091771, 1092395, 1100186, 1100188, 1100189.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1439-2017
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.