Z-1441-2022 Class II Ongoing
FDA device recall Z-1441-2022 was initiated by Becton Dickinson & Company on May 9, 2022 and is designated Class II. Reason for recall: May have the potential for leakage at the housing component of the stopcock, may result in delay or interruption in treatment, exposure to infusate and biohazardous material, under dosing/under… The recall status is ongoing. Affected quantity: 44,040.
Recall Details
- Product Type
- Devices
- Report Date
- Jul 27, 2022
- Initiation Date
- May 9, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 44,040
Product Description
Nexiva with Single BD Connecta Stopcock Version (20 Ga Blue) Catalog No.383682 (OUS)
Reason for Recall
May have the potential for leakage at the housing component of the stopcock, may result in delay or interruption in treatment, exposure to infusate and biohazardous material, under dosing/under infusion, contamination and/or air ingress.
Distribution Pattern
Worldwide distribution - US Nationwide distribution in the states of AK, AR, AZ, CA, CO, FL, GA, IL, IN, KS, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OR, PA, SD, TN, TX, UT, VA, WA, WV and the countries of Belgium, China, India, Indonesia, Japan, Korea, Nepal, Philippines, Singapore, Sri Lanka, Thailand, Vietnam.
Code Information
GTIN: 382903836826 Lot Numbers/Exp.Date: 1082423 3/31/2024 1117627 4/30/2024 1140859 5/31/2024 1210968 7/31/2024