Z-1449-2018 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-1449-2018 was initiated by Siemens Medical Solutions USA, Inc on Feb 15, 2018 and is designated Class II. Reason for recall: to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on… The recall status is terminated (terminated Apr 8, 2020). Affected quantity: 1301.

Recall Details

Product Type
Devices
Report Date
May 9, 2018
Initiation Date
Feb 15, 2018
Termination Date
Apr 8, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1301

Product Description

SOMATOM Scope Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Reason for Recall

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Distribution Pattern

USA (nationally) Distribution.

Code Information

10046799

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1449-2018