Z-1457-2024 Class II Ongoing

Recalled by Covidien, LP — North Haven, CT

FDA device recall Z-1457-2024 was initiated by Covidien, LP on Feb 14, 2024 and is designated Class II. Reason for recall: Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads… The recall status is ongoing. Affected quantity: 30 Units.

Recall Details

Product Type
Devices
Report Date
Apr 10, 2024
Initiation Date
Feb 14, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
30 Units

Product Description

Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT

Reason for Recall

Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of France, Hong Kong, Israel and Japan.

Code Information

Model Number: SIGSDS30CTVT; UDI/DI: 10884521741881; Lot number: N3L1909UY;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1457-2024
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.