Z-1461-2018 Class II Terminated

Recalled by Cook Inc. — Bloomington, IN

Recall Details

Product Type
Devices
Report Date
April 25, 2018
Initiation Date
March 22, 2018
Termination Date
May 13, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,267

Product Description

Cook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN G49275. Intended for the aspiration of body fluids and cells, in particular oocyte aspiration.

Reason for Recall

Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.

Distribution Pattern

Worldwide Distribution -- USA, Argentina, Bolivia, Brazil, Canada, Colombia, Guatemala, Mexico, Panama, Peru, Uruguay, and Venezuela.

Code Information

All lots