Z-1481-2023 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-1481-2023 was initiated by Siemens Medical Solutions USA, Inc on Nov 23, 2022 and is designated Class II. Reason for recall: Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient… The recall status is ongoing. Affected quantity: 704 units in the United States (1945 units worldwide).

Recall Details

Product Type
Devices
Report Date
May 31, 2023
Initiation Date
Nov 23, 2022
Voluntary/Mandated
FDA Mandated
Product Quantity
704 units in the United States (1945 units worldwide)

Product Description

LUMINOS Lotus Max (VF11)

Reason for Recall

Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube.

Distribution Pattern

US Nationwide - Worldwide Distribution

Code Information

Model: 11574100