Z-1483-2024 Class II Ongoing

Recalled by Microbiologics Inc — Saint Cloud, MN

FDA device recall Z-1483-2024 was initiated by Microbiologics Inc on March 6, 2024 and is designated Class II. Reason for recall: The positive control material (Rhizopus stolonifer) within the KWIK-STIK assembly was contaminated with another fungal organism (Aspergillus brasiliensis). As a result, users may observe A. brasiliens… The recall status is ongoing. Affected quantity: 19 units.

Recall Details

Product Type
Devices
Report Date
April 17, 2024
Initiation Date
March 6, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
19 units

Product Description

KWIK-STIK Rhizopus stolonifer (+) derived from ATCC 6227b, a) REF 0209P and b) REF 0209K, Quality control kit for culture media

Reason for Recall

The positive control material (Rhizopus stolonifer) within the KWIK-STIK assembly was contaminated with another fungal organism (Aspergillus brasiliensis). As a result, users may observe A. brasiliensis growth when using the control material.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of CA, MI, KY, MO, NM, TN, PR, and the countries of France, Columbia, Kazakhstan, Peru, Puerto Rico, South Africa.

Code Information

a) REF 0209P, UDI/DI 20845357003113, Lot Numbers: 209-37-1, 209-37-3, 209-37-4, 209-37-5; b) REF 0209K, UDI/DI 30845357003127, Lot Numbers: 209-37-2