Z-1454-2025 Class II Ongoing

Recalled by Microbiologics Inc — Saint Cloud, MN

FDA device recall Z-1454-2025 was initiated by Microbiologics Inc on Feb 20, 2025 and is designated Class II. Reason for recall: The RSV target may give a late Ct value and could potentially not pass QC. The recall status is ongoing. Affected quantity: 28 units.

Recall details

Product Type
Devices
Report Date
Apr 2, 2025
Initiation Date
Feb 20, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
28 units

Product description

Helix Elite Inactivated Standard: Inactivated Influenza A/B and Respiratory Syncytial Virus, Multi-analyte, unassayed quality control material, Catalog Number HE0044N

Reason for recall

The RSV target may give a late Ct value and could potentially not pass QC.

Distribution pattern

Worldwide - US Nationwide distribution in the states of KY, MT, NJ, NY, WI and the countries of Ireland, Germany, UK.

Code information

UDI-DI: 10845357042184; Lot Number: HE0044-131