Z-1163-2025 Class II Ongoing

Recalled by Microbiologics Inc — Saint Cloud, MN

Recall Details

Product Type
Devices
Report Date
February 19, 2025
Initiation Date
January 20, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 units

Product Description

Microbiologics Multi-Strain Human Papillomavirus (HPV) Control Panel (Inactivated Pellet), REF 8234; in-vitro diagnostic test

Reason for Recall

Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Netherlands, Sweden, UK, France, Australia, New Zealand.

Code Information

UDI/DI 70845357043077, Lot Number 8234-22