Z-1163-2025 Class II Ongoing

Recalled by Microbiologics Inc — Saint Cloud, MN

FDA device recall Z-1163-2025 was initiated by Microbiologics Inc on Jan 20, 2025 and is designated Class II. Reason for recall: Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded… The recall status is ongoing. Affected quantity: 16 units.

Recall details

Product Type
Devices
Report Date
Feb 19, 2025
Initiation Date
Jan 20, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 units

Product description

Microbiologics Multi-Strain Human Papillomavirus (HPV) Control Panel (Inactivated Pellet), REF 8234; in-vitro diagnostic test

Reason for recall

Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Netherlands, Sweden, UK, France, Australia, New Zealand.

Code information

UDI/DI 70845357043077, Lot Number 8234-22

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1163-2025
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1163-2025. Report a data issue: hello@regdatalab.com.