FDA device recall Z-1494-2020 was initiated by Philips North America, LLC on Dec 30, 2019 and is designated Class II. Reason for recall: A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible. The recall status is terminated (terminated Nov 17, 2022). Affected quantity: 87.
Recall details
- Product Type
- Devices
- Report Date
- Mar 25, 2020
- Initiation Date
- Dec 30, 2019
- Termination Date
- Nov 17, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 87
Product description
Reason for recall
Distribution pattern
Nationwide domestic distribution. Foreign distribution worldwide.