Z-1317-2026 Class II Ongoing

Recalled by Philips North America Llc — Cambridge, MA

FDA device recall Z-1317-2026 was initiated by Philips North America Llc on Jan 7, 2026 and is designated Class II. Reason for recall: As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If… The recall status is ongoing. Affected quantity: 283 units.

Recall details

Product Type
Devices
Report Date
Feb 18, 2026
Initiation Date
Jan 7, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
283 units

Product description

Spectral CT; Product Code (REF): 728333;

Reason for recall

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Macao, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, R¿union, Russian Fed., Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkmenistan, United Kingdom, Uzbekistan, Vietnam.

Code information

Product Code (REF): 728333; UDI-DI: 00884838101111; Serial Numbers: 10055, 10075, 10086, 10096, 10106, 10116, 10126, 10136, 10146, 10156, 10166, 10176, 10186, 10196, 10206, 10217, 10227, 10237, 10247, 10257, 10267, 10277, 10289, 10302, 10314, 12009, 12019, 12029, 396013, 10056, 10076, 10087, 10097, 10107, 10117, 10127, 10137, 10147, 10157, 10167, 10177, 10187, 10197, 10207, 10218, 10228, 10238, 10248, 10258, 10268, 10278, 10290, 10303, 12000, 12010, 12020, 12030, 398003, 10057, 10077, 10088, 10098, 10108, 10118, 10128, 10138, 10148, 10158, 10168, 10178, 10188, 10198, 10208, 10219, 10229, 10239, 10249, 10259, 10269, 10279, 10291, 10304, 12001, 12011, 12021, 12031, 398004, 10058, 10078, 10089, 10099, 10109, 10119, 10129, 10139, 10149, 10159, 10169, 10179, 10189, 10199, 10209, 10220, 10230, 10240, 10250, 10260, 10270, 10280, 10292, 10305, 12002, 12012, 12022, 396005, 10059, 10079, 10090, 10100, 10110, 10120, 10130, 10140, 10150, 10160, 10170, 10180, 10190, 10200, 10210, 10221, 10231, 10241, 10251, 10261, 10271, 10281, 10293, 10306, 12003, 12013, 12023, 396006, 10060, 10080, 10091, 10101, 10111, 10121, 10131, 10141, 10151, 10161, 10171, 10181, 10191, 10201, 10211, 10222, 10232, 10242, 10252, 10262, 10272, 10284, 10294, 10307, 12004, 12014, 12024, 396008, 10061, 10081, 10092, 10102, 10112, 10122, 10132, 10142, 10152, 10162, 10172, 10182, 10192, 10202, 10212, 10223, 10233, 10243, 10253, 10263, 10273, 10285, 10298, 10308, 12005, 12015, 12025, 396009, 10062, 10082, 10093, 10103, 10113, 10123, 10133, 10143, 10153, 10163, 10173, 10183, 10193, 10203, 10213, 10224, 10234, 10244, 10254, 10264, 10274, 10286, 10299, 10310, 12006, 12016, 12026, 396010, 10070, 10083, 10094, 10104, 10114, 10124, 10134, 10144, 10154, 10164, 10174, 10184, 10194, 10204, 10214, 10225, 10235, 10245, 10255, 10265, 10275, 10287, 10300, 10311, 12007, 12017, 12027, 396011, 10074, 10085, 10095, 10105, 10115, 10125, 10135, 10145, 10155, 10165, 10175, 10185, 10195, 10205, 10215, 10226, 10236, 10246, 10256, 10266, 10276, 10288, 10301, 10312, 12008, 12018, 12028, 396012,

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1317-2026
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1317-2026. Report a data issue: hello@regdatalab.com.