Z-1532-2025 Class II Ongoing

Recalled by Omnicell, Inc. — Cranberry Township, PA

Recall Details

Product Type
Devices
Report Date
April 16, 2025
Initiation Date
February 20, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3

Product Description

IVX Station Reconstitution Transfer Set. Model Number: 298921215. Packaging: 1 set per White Tyvek Pouch, 5 Tyvek Pouches per shelf box, 10 shelf boxes per case box. IVX Station Reconstitution Transfer Set is used for the reconstitution of drug vials in hospital pharmacies. It is indicated for use with the IVX Station Pharmacy Compounding System.

Reason for Recall

IVX Station Reconstitution Transfer Sets were shipped in advance of the IVX Station upgrade and not intended for clinical use but are engineering devices intended for use during set up and system calibration. The items were not shipped in a validated packing configuration and sterility cannot be ensured.

Distribution Pattern

Distributed in the US to Alabama

Code Information

Model number: 298921215. UDI Number: 0086000726042101124301840119041525, 0086000726042101124300840119041525, 0086000726042101124301680119041525. Lot number 011243.