Z-1617-2023 Class II Ongoing
FDA device recall Z-1617-2023 was initiated by OmniCell, Inc. on Mar 27, 2023 and is designated Class II. Reason for recall: IVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling The recall status is ongoing. Affected quantity: 6 units.
Recall details
- Product Type
- Devices
- Report Date
- May 31, 2023
- Initiation Date
- Mar 27, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6 units
Product description
Omnicell IVX Station-For use in the automated preparation of pharmaceutical admixtures and syringes Model Number: 298921373
Reason for recall
IVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling
Distribution pattern
AL, NY
Code information
UDI-DI: 00860006285005 Serial Numbers: IVXS000014, IVXS000015, IVXS000016, IVXS000017, IVXS000018. IVXS000019
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1617-2023
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
Need the full FDA record for OmniCell, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.