Z-1617-2023 Class II Ongoing

Recalled by OmniCell, Inc. — Warrendale, PA

FDA device recall Z-1617-2023 was initiated by OmniCell, Inc. on Mar 27, 2023 and is designated Class II. Reason for recall: IVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling The recall status is ongoing. Affected quantity: 6 units.

Recall details

Product Type
Devices
Report Date
May 31, 2023
Initiation Date
Mar 27, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 units

Product description

Omnicell IVX Station-For use in the automated preparation of pharmaceutical admixtures and syringes Model Number: 298921373

Reason for recall

IVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling

Distribution pattern

AL, NY

Code information

UDI-DI: 00860006285005 Serial Numbers: IVXS000014, IVXS000015, IVXS000016, IVXS000017, IVXS000018. IVXS000019

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1617-2023
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for OmniCell, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1617-2023. Report a data issue: hello@regdatalab.com.