Z-1544-2018 Class III Terminated

Recalled by Becton Dickinson & Co. — Sparks, MD

Recall Details

Product Type
Devices
Report Date
May 9, 2018
Initiation Date
March 2, 2018
Termination Date
July 2, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13,898

Product Description

BD MAX DNA MMK (SPC) For Laboratory Use, catalog no. 442829

Reason for Recall

The package insert incorrectly lists the volume of the primer and probe diluent to use for the reconstitution of the master mix tube. Adding the incorrect amount of diluent could have an effect on the assay.

Distribution Pattern

Distributed to accounts in 8 states: CA, FL, NJ, NY, OR, PA, VA, and WI. International distribution to Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Europe, Hong Kong, Japan, Singapore, Taiwan, and Thailand.

Code Information

All since 08/20/2014