Z-1546-2015 Class II Terminated

Recalled by Pega Medical Inc. — Laval

Recall Details

Product Type
Devices
Report Date
May 6, 2015
Initiation Date
April 10, 2015
Termination Date
June 29, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9 units (5 units in the US and 4 units international)

Product Description

Female Driver, Part number FDr101, instrument used for the insertion of the Fassier-Duval IM Telescopic System. For orthopedic procedures. This instrument is included in the Fassier-Duval IM Telescopic System instrument tray. This instrument is marked with the catalogue number and lot number.

Reason for Recall

The hexagonal tip of the Female Driver instrument for the Fassier-Duval IM Telescopic System is 0.06 mm under the specifications. This may result in a reduction of the total strength of the instrument, possibly leading to deformation, wear, and tip breakage.

Distribution Pattern

Worldwide distribution. US in Delaware, Florida, Nebraska, and New York; Australia, Brazil, and United Kingdom.

Code Information

Lot 290609-01