Z-1848-2015 Class II Terminated
FDA device recall Z-1848-2015 was initiated by Pega Medical Inc. on Jun 15, 2015 and is designated Class II. Reason for recall: SCFE Drivers of the Free-Gliding SCFE Screw System are recalled because the handle of the instruments may detach from the shaft if high torques are applied in counter-clockwise rotation. The recall status is terminated (terminated Dec 29, 2015). Affected quantity: 4 units.
Recall details
- Product Type
- Devices
- Report Date
- Jul 8, 2015
- Initiation Date
- Jun 15, 2015
- Termination Date
- Dec 29, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4 units
Product description
Slipped Capital Femoral Epiphysis (SCFE) Drivers of the Free-Gliding SCFE Screw System with catalogue number SCF-MLD265 and SCF-MLD273.
Reason for recall
SCFE Drivers of the Free-Gliding SCFE Screw System are recalled because the handle of the instruments may detach from the shaft if high torques are applied in counter-clockwise rotation.
Distribution pattern
US Distribution to the states of : FL and TN.
Code information
Part Number: SCF-MLD265, LOT#: 130703-01. Part Number: SCF-MLD273, LOT#: 130703-02.