Z-1547-2013 Class II Terminated

Recalled by Biomet, Inc. — Warsaw, IN

FDA device recall Z-1547-2013 was initiated by Biomet, Inc. on Apr 29, 2013 and is designated Class II. Reason for recall: Biomet Trauma ("Biomet") has initiated a recall of Retrograde Femoral Connecting Bolt, which involves Part Number 14-442021. The connecting bolt has an undersized diameter specification that may… The recall status is terminated (terminated May 6, 2015). Affected quantity: 75.

Recall Details

Product Type
Devices
Report Date
Jun 26, 2013
Initiation Date
Apr 29, 2013
Termination Date
May 6, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
75

Product Description

Biomet Trauma Phoenix Retrograde Femoral Connecting Bolt Trauma Fixation Systems. Product Usage: The Phoenix Retrograde Femoral Nail is indicated for alignment, stabilization, and fixation of fractures caused by trauma or disease and the fixation of femurs that have been surgically prepared (osteotomy) for correction of deformity.

Reason for Recall

Biomet Trauma ("Biomet") has initiated a recall of Retrograde Femoral Connecting Bolt, which involves Part Number 14-442021. The connecting bolt has an undersized diameter specification that may cause an interference fit with the 4mm hex driver (Part Number: 41024) near the edge of the tolerance. If the 4mm hex driver becomes stuck in the connecting bolt it cannot be taken apart and a delay in su

Distribution Pattern

Worldwide Distribution - USA Nationwide: CA, TX, NY, PA, NJ, FL, OH, MI, CO, SD, IN, MT, GA, and WI. and the countries of Netherlands, Costa Rica, Japan, and San Juan.

Code Information

Catalog number: 14-442021 and lot number:757910

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1547-2013
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.