Z-1564-2016 Class II Terminated

Recalled by Hansen Medical Inc — Mountain View, CA

Recall Details

Product Type
Devices
Report Date
May 4, 2016
Initiation Date
March 25, 2016
Termination Date
May 6, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8 kits

Product Description

Magellan Robotic Catheter Accessory Kits, Model number MACC 50 and MACC80. Cardiology: The Hansen Medical Magellan System is a robotic control system that enables positioning and visualization of a Magellan Catheter and an off-the-shelf guide wire while allowing a physician to remain seated and away from the x-ray radiation source. The Accessory Kits contain disposable components that keep the Magellan Catheters and 3rd-party guide wires performing smoothly during system operation.

Reason for Recall

Firm's quarterly dose audit testing showed that one batch was radiated below the augmented dose and therefore did not meet the augmented dose of 28.4 kGy.

Distribution Pattern

Nationwide Distribution to OH, TX. FL, CA and GA.

Code Information

Lot number 20151117