Z-1564-2016 Class II Terminated

Recalled by Hansen Medical Inc — Mountain View, CA

FDA device recall Z-1564-2016 was initiated by Hansen Medical Inc on Mar 25, 2016 and is designated Class II. Reason for recall: Firm's quarterly dose audit testing showed that one batch was radiated below the augmented dose and therefore did not meet the augmented dose of 28.4 kGy. The recall status is terminated (terminated May 6, 2016). Affected quantity: 8 kits.

Recall details

Product Type
Devices
Report Date
May 4, 2016
Initiation Date
Mar 25, 2016
Termination Date
May 6, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8 kits

Product description

Magellan Robotic Catheter Accessory Kits, Model number MACC 50 and MACC80. Cardiology: The Hansen Medical Magellan System is a robotic control system that enables positioning and visualization of a Magellan Catheter and an off-the-shelf guide wire while allowing a physician to remain seated and away from the x-ray radiation source. The Accessory Kits contain disposable components that keep the Magellan Catheters and 3rd-party guide wires performing smoothly during system operation.

Reason for recall

Firm's quarterly dose audit testing showed that one batch was radiated below the augmented dose and therefore did not meet the augmented dose of 28.4 kGy.

Distribution pattern

Nationwide Distribution to OH, TX. FL, CA and GA.

Code information

Lot number 20151117

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1564-2016
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Hansen Medical Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1564-2016. Report a data issue: hello@regdatalab.com.