Z-0896-2015 Class II Terminated

Recalled by Hansen Medical Inc — Mountain View, CA

Recall Details

Product Type
Devices
Report Date
January 7, 2015
Initiation Date
December 4, 2014
Termination Date
September 3, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
59 USA, 40 OUS

Product Description

Sensei X Robotic Catheter System, model numbers 02057 and 08257. - Set up joint (SUJ)mechanical clamp that secures the SUJ to the catheter laboratory table. Product Usage: The Sensei X Robotic Catheter System is an electromechanical software controlled system designed to facilitate navigation and control of Hansen Medicals robotically steerable catheters for collecting electrophysiological data within the heart atria with electro-anatomic mapping and recording systems.

Reason for Recall

The Set Up Joint subassembly of the Sensei X Robotic Catheter System may become loose relative to the operating table with the potential for the device to fall causing, uncontrolled catheter movement. cardiac tampenade, delayed procedure, or crush injury to the patient and/or user.

Distribution Pattern

Worldwide Distribution - US Nationwide and the countries of Italy, Czech Republic, Russia, Spain, Singapore, France, Germany, Denmark, Netherlands and Australia.

Code Information

Model numbers 02057 and 08257, all installed Sensei X Robotic systems are subject to this action.