Z-0896-2015 Class II Terminated

Recalled by Hansen Medical Inc — Mountain View, CA

FDA device recall Z-0896-2015 was initiated by Hansen Medical Inc on Dec 4, 2014 and is designated Class II. Reason for recall: The Set Up Joint subassembly of the Sensei X Robotic Catheter System may become loose relative to the operating table with the potential for the device to fall causing, uncontrolled catheter movement… The recall status is terminated (terminated Sep 3, 2015). Affected quantity: 59 USA, 40 OUS.

Recall details

Product Type
Devices
Report Date
Jan 7, 2015
Initiation Date
Dec 4, 2014
Termination Date
Sep 3, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
59 USA, 40 OUS

Product description

Sensei X Robotic Catheter System, model numbers 02057 and 08257. - Set up joint (SUJ)mechanical clamp that secures the SUJ to the catheter laboratory table. Product Usage: The Sensei X Robotic Catheter System is an electromechanical software controlled system designed to facilitate navigation and control of Hansen Medicals robotically steerable catheters for collecting electrophysiological data within the heart atria with electro-anatomic mapping and recording systems.

Reason for recall

The Set Up Joint subassembly of the Sensei X Robotic Catheter System may become loose relative to the operating table with the potential for the device to fall causing, uncontrolled catheter movement. cardiac tampenade, delayed procedure, or crush injury to the patient and/or user.

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of Italy, Czech Republic, Russia, Spain, Singapore, France, Germany, Denmark, Netherlands and Australia.

Code information

Model numbers 02057 and 08257, all installed Sensei X Robotic systems are subject to this action.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0896-2015
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Hansen Medical Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0896-2015. Report a data issue: hello@regdatalab.com.