Z-1568-2014 Class II Terminated

Recalled by Medtronic Neuromodulation — Minneapolis, MN

FDA device recall Z-1568-2014 was initiated by Medtronic Neuromodulation on Apr 4, 2014 and is designated Class II. Reason for recall: Some refill kits in the affected lots may contain an extension tubing set that is occluded and healthcare professionals may be unable to aspirate or fill the SynchroMed pumps. There is no problem… The recall status is terminated (terminated Nov 5, 2014). Affected quantity: 13,784.

Recall Details

Product Type
Devices
Report Date
May 14, 2014
Initiation Date
Apr 4, 2014
Termination Date
Nov 5, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13,784

Product Description

Refill Kit 8551/856X (includes Models 8561, 8562 and 8564) for use with Medtronic Implantable Programmable Infusion Pumps. Product Usage: The Refill Kits are intended for use in refilling Medtronic programmable, implantable infusion pumps with the exception of Medtronic MiniMed infusion pumps.

Reason for Recall

Some refill kits in the affected lots may contain an extension tubing set that is occluded and healthcare professionals may be unable to aspirate or fill the SynchroMed pumps. There is no problem with the Lioresal Intrathecal (baclofen injection) contained in the affected drug refill kits. There have been no patient injuries beyond additional needle sticks associated with this issue.

Distribution Pattern

Worldwide Distribution - USA Nationwide in the states of NM, ND, RI, and VT and the countries of Australia, Singapore, Thailand, Central and Eastern Europe, Western Europe, Middle East, & Panama.

Code Information

Lot numbers CS3643, CS3613, CS3663, N441998, N442081, N442131, N440259, N440527, N440582, N440654, N440909, N440978, N441199, N441268, N441806, N441817, N442178, N442513, N442532, N442560, N442588, N442831, and N443427. Affected lots were distributed between January 2014 and March 2014.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1568-2014
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.