Z-1570-2014 Class II Terminated

Recalled by Medtronic Neuromodulation — Minneapolis, MN

FDA device recall Z-1570-2014 was initiated by Medtronic Neuromodulation on Feb 26, 2014 and is designated Class II. Reason for recall: This recall provides important new information regarding overinfusion associated with the Medtronic SynchroMed II Implantable Pump. Overinfusion can result in a life-threatening overdose and can also… The recall status is terminated (terminated Sep 28, 2018). Affected quantity: 195,198 pumps (146,435 US, 48,763 OUS).

Recall Details

Product Type
Devices
Report Date
May 14, 2014
Initiation Date
Feb 26, 2014
Termination Date
Sep 28, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
195,198 pumps (146,435 US, 48,763 OUS)

Product Description

Medtronic SynchroMed¿ II Implantable Drug Infusion Pump, Model 8637-20, 8637-40. The implantable Medtronic SynchroMed II programmable pumps are part of an infusion system that stores and delivers a prescribed drug to a specific site. The implanted infusion system consists of a SynchroMed II pump and a catheter.

Reason for Recall

This recall provides important new information regarding overinfusion associated with the Medtronic SynchroMed II Implantable Pump. Overinfusion can result in a life-threatening overdose and can also result in drug withdrawal due to premature emptying of the pump. Due to the low reported rate of occurrence of this issue and the inability to predict which pumps may be at risk, Medtronic is not re

Distribution Pattern

Worldwide Distribution - All states in USA. OUS: List not provided at this time.

Code Information

This Medical Device Correction notification affects all SynchroMed II pumps.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1570-2014
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.