Z-1580-2013 Class I Terminated

Recalled by Medtronic Neuromodulation — Minneapolis, MN

FDA device recall Z-1580-2013 was initiated by Medtronic Neuromodulation on Jun 3, 2013 and is designated Class I. Reason for recall: Medtronic Neuromodulation is providing Healthcare Providers with safety information and patient management recommendations related to the potential for electrical shorting internal to they SynchroMed… The recall status is terminated (terminated Nov 16, 2015). Affected quantity: 261,109 total devices estimated implanted Worldwide.

Recall Details

Product Type
Devices
Report Date
Jul 3, 2013
Initiation Date
Jun 3, 2013
Termination Date
Nov 16, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
261,109 total devices estimated implanted Worldwide

Product Description

Medtronic SynchroMed EL Programmable pumps, Models 8626, 8626L, 8627, 8627L, (The SynchroMed EL Pumps are supplied in 10 ml or 18 ml reservoir size.) Sterilized using ethylene oxide. The pump is part of an infusion system that stores and delivers a prescribed drug to a specific site. The drug is stored in the pump reservoir. Per a programmed prescription, the drug moves from the pump reservoir, through the pump tubing, catheter port, and catheter, to the infusion site. The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the following drugs or fluids:

Reason for Recall

Medtronic Neuromodulation is providing Healthcare Providers with safety information and patient management recommendations related to the potential for electrical shorting internal to they SynchroMed infusion pump. An electrical short circuit in a feedthrough may present as a motor stall or low battery reset/alarm and lead to a loss of or reduction in therapy which may result in the return of und

Distribution Pattern

worldwide

Code Information

all SynchroMed EL pumps

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1580-2013
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.