Z-1583-2022 Class II Ongoing

Recalled by Mckesson Medical-Surgical Inc. Corporate Office — Richmond, VA

Recall Details

Product Type
Devices
Report Date
August 31, 2022
Initiation Date
May 25, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
a. 27 boxes b. 1 box

Product Description

a. Rapid Test Kit Coag-Sense Professional Blood Coagulation Test Prothrombin Time Test (PT/INR) Whole Blood Sample 50 Tests-IVD. Part No. 0356-50. b. Coagulation Control Kit Coag-Sense PT / INR Test (Prothrombin Time Test) Low Level / High Level 10 Strips Part No. 03P69-10.

Reason for Recall

Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

Distribution Pattern

US Nationwide

Code Information

a. GTIN: 00861900000336; 10861900000395; 00861900000398. Lot codes distributed between 5/30/2021 - 8/24/2021. b. GTIN 00861900000343 Lots Distributed: 8/20/2021 - 8/20/2021