Z-1586-2022 Class II Ongoing

Recalled by Mckesson Medical-Surgical Inc. Corporate Office — Richmond, VA

Recall Details

Product Type
Devices
Report Date
August 31, 2022
Initiation Date
May 25, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
a. 28 b. 22

Product Description

a. ABX PENTRA CREATININE 120CP JAFFE 120/130. Ref: A11A01933. b. Reagent General Chemistry Creatinine Model Number: 1220001933

Reason for Recall

Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

Distribution Pattern

US Nationwide

Code Information

GTIN: 03610230005736 Lots Distributed: 5/30/2021 - 9/8/2021