Z-1590-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 9, 2018
- Initiation Date
- January 31, 2018
- Termination Date
- April 17, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 751
Product Description
StoneBreaker Pneumatic Lithotripter, Catalog number: SBL-KIT1; GPN: G52604 UDI:(01)00827002526044
Reason for Recall
The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
Distribution Pattern
US Nationwide, OUS countries: AE, AT, AU, BE, BG, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DJ, DK, DO, EC, EE, EG, ES, FI, FJ, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IT, JO, KZ, LB, LT, LV, MA, MO, MX, MY, NI, NL, NO, NZ, PE, PH, PK, PL, PR, PT, QA, RO, RU, SA, SE, SG, SI, SK, SV, TH, TT, TW, UA, UY, XK, ZA
Code Information
All Lots