Z-1592-2026 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- April 1, 2026
- Initiation Date
- January 23, 2026
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 220 units
Product Description
Volar Radius Plate with Angular Stability with the following description: Radius Plate PROlock II. Article Numbers: 21031-3, 21031-5, 21032-11, 21032-3, 21032-5, 21033-3, 21033-5, 21034-3, 21034-5, 21035-3, 21036-3.
Reason for Recall
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Distribution Pattern
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Code Information
Article Number (Lot Numbers): 21031-3 (679/042207, 679/032240, 679/0323115, 679/052317, 26/F70606, 679/032253, 643/032501, 643/072503, 643/102501), 21031-5 (43/0015, 643/102201, 643/102201, 679/0323121, 679/0323120, 679/0323120, 679/0323120, 679/0323120, 679/0323120, 679/0323120, 679/0323120, 679/0323120, 643/102201, 643/102201, 643/102201, 643/102201, 679/0323120, 679/0323120), 21032-11 (679/112312), 21032-3 (679/032241, 643/0623, 679/042293, 679/0323122, 679/062367, 679/052318, 643/022501, 643/062501, 643/072501, 643/102502), 21032-5 (679/122136, 679/0323143, 679/042337), 21033-3 (679/062212, 679/022301, 679/042338, 679/022302), 21033-5 (643/0629, 26/F24901, 679/012311), 21034-3 (679/062221, 679/012312, 679/072217, 679/042339), 21034-5 (643/102202, 679/012314, 679/012313), 21035-3 (679/062222, 679/042308, 643/012501, 643/0548), 21036-3 (679/062213, 679/042309, 679/112317),